Timing Reference Outputs ======================== UUID: ``E935EEA3-4CC8-4BD6-B128-2B72C10D9352`` Expected outputs: * fM measured at 2 MHz +/- 0.1% (1.998-2.002 MHz); * SH measured against its 10 microsecond pulse target and checked for alignment that preserves the required ICG off time; * ICG measured against an 8 ms period and 7.388 ms pulse targets; * ADC-trigger behavior observed during the capture application boundary, which is ``Not In Docs`` until the shared control contract is specified; and * a pass/fail record linking each capture to the hardware and requirement revisions. The ADC-trigger frequency and all tolerances beyond the stated nominal values remain ``Not In Docs``.